The numerous control systems make X-Mind DC very safe to use: self-diagnosis of control panel elements each time it is started up, check of the electric installation parameters by the timer and automatic compensation of the exposure time depending on variations in mains voltage ensure very high quality performance.
 The X-Mind range pantograph arms provide great flexibility of use and accurate positioning. The new, light alloy mechanism has smooth contours and protective casing at the joints.
 The new X-Mind DC generator is fitted with a Toshiba DG-073B-DC double anode tungsten tube (intensity of 4mA and 8mA). It generates a constant voltage of 60kV or 70kV. It operates at high frequencies and constant potential which enables very high quality radiation to be emitted, whatever the conditions of use. The choice of voltage (60 or 70kV) and intensity (4 or 8mA) enables optimal diagnosis of intraoral X-rays with a 35% reduction in exposure time, compared with monophased units. The timer can control up to two DC generators. The extremely short exposure times (minimum 0.02 sec.) are suitable for digital radiology with CCD, CMOS or phosphor plate sensors (radiation is 50% less that that used with “E” type films).
 Three extension arms are available (wall mounting): 41cm (16.2"), 82.5cm (32.5") and 110cm (43.5") giving reaches respectively of 143cm, 184cm and 212cm (with a long cone).
 X-Mind timer: microprocessor-controlled, exposure time can be customised and programmed; one key allows you to switch immediately from conventional intra-oral X-ray to digital X-ray. Pre-programmed keys for special pictures
 Classification: Electromedical equipment, Class 1 type B
 Supply voltage: 230V- monophase 50/60Hz
 Power absorption at 230V: 1.4kVA
 Resistance of the line: 0.5 ohms
 X-ray tube: New Toshiba DG 073B DC tube
 High voltage: 60-70kV
 Anode current: 4-8mA
 Focal spot: 0.7mm
 Total filtration: Equivalent to 2mm Al at 70kV
 Tube inherent filtration: Equivalent to 0.8mm Al at 70kV
 Leakage radiation: Less than 0.25mGy/hour maximum at a distance of 1m from focal spot
 Long cone: Focus to cylinder tip distance = 31cm (12”)
 Timer: Exposure time can be set from 0.02 to 3.2 seconds
 X-ray emission control: “Dead man” switch with 3m spiral cable
 Total weight: 25kg (weight of the head: 5.5kg)
 
 
  |